基于微磁理论的镀层工件裂纹检测仪,检测仪性能特点

发布时间:2022-12-29 07:30:01 人气:

摘要:以具有镀层的铁磁零部件的裂纹检测为背景,阐述了微磁裂纹检测仪的主要试验方法、裂纹特征二次梯度提取及裂纹检测的实现。
关键词:镀层零件;微磁检测;二次梯度

Instrument of Testing Crack for the Parts with Plating Layer Based on Micromagnetic Theory
CHEN Guiqin1, XU Zhangsui2, WANG Feng2, ZHOU Hailin2
(1.Hebei Higher Special School of Engineering, Cangzhou 0 61001, China;
2.Ordnance Engineering College, Shijiazhuang 050003, China)
Abstract: Using the crack detection of ferromagnetic materials p arts with plating layer as a sample, the micromagnetic crack testing principle s, the crack characteristic extracting based on double gradient, and realizing o f crack testing were introduced in this paper.
Key words: parts with plating layer; micromagnetic detection; double gradient

镀层工件是指在钢材料表面镀有如铬、镍、锌等金属的工件,镀层厚度约为20~80μm。这些金属的磁导率近于1,不适于采用磁粉探伤;其镀层覆盖在铁磁工件表面,也不适于电涡流探伤和超声探伤方法。笔者在国家自然科学基金资助项目研究中,运用微磁检测技术对这种工件裂纹进行了检测,取得了满意的效果。

1 裂纹微磁检测试验方法

由材料的微观磁特性可知:(1)当外加机械应力时,会使材料中的晶格组织发生变化。在损伤区边缘出现附加磁极,产生磁荷聚集,形成磁场,材料对外显示磁性。缺陷不再扩展,其缺陷磁场强度保持不变。(2)当材料内部存在裂纹、夹杂等缺陷时,也会破坏原来的晶格,形成累计磁场,对外显磁性。

因此在内应力集中或缺陷集中的地方,金属导磁率下降,形成一内部磁源。这一内部磁源向金属表面传递,形成泄漏磁场,其切向分量*大,法向分量具有从正过零到负的变化过程。

金属构件缺陷的存在,就一定产生磁畴固定结点,形成内部磁场。这种磁场十分微弱,按微磁学试验方法称为微磁点,其检测过程称为微磁检测。

2 裂纹磁畴结点磁荷变异